ISO 15189:2022 internal audit: how to plan and conduct one in a medical laboratory
Healthcare Safety and Quality Institute, 28 November 2026. 3 min read.
ISO 15189:2022 requires a medical laboratory to carry out internal audits at planned intervals. The requirement sits in clause 8.8.3. The purpose is to learn whether the management system meets the standard and the laboratory's own requirements, and whether it is working in practice. This guide shows how to do it.
Start with the annual programme
The audit programme is the plan for the year. It should cover:
- every clause that applies, from 4 (general requirements) to 8 (management system requirements),
- every section of the laboratory, and all three phases: pre-examination, examination and post-examination,
- sample collection centres and point-of-care testing, where the laboratory is responsible for them.
Plan with risk in mind. An area with new equipment, new staff, past findings or complaints deserves to be audited more often than a stable one.
Choose auditors who are competent and independent
An auditor should be trained in the standard and in audit method, and should not audit their own work. In a small laboratory, staff of one section can audit another. The quality manager's own activities should be audited by someone else.
Three ways to audit
- A horizontal audit takes one requirement, such as equipment calibration or staff competence, and checks it across all sections.
- A vertical audit takes one patient sample and follows it from the request form to the released report, checking every record on the way: collection, transport, acceptance, the analyser, the reagent lot, the calibration, the internal quality control of that day, the person who did the test and their competence record, the review and the report.
- A witness audit means watching a person carry out a task, such as a phlebotomy or an examination, against the documented procedure.
A good programme uses all three. The vertical audit is the quickest way to find out whether the system really holds together.
Prepare a checklist, but do not read from it
A clause-wise checklist makes sure nothing is missed. Write each point as a question that needs evidence, not a yes or no. Ask "show me", then look at records, watch the work and talk to the person doing it.
Write findings that can be acted on
A nonconformity should state three things: the requirement, what was found, and the evidence. For example: the procedure requires internal quality control at two levels on each day of testing; no control was run for the glucose assay on two named dates; the control log and the patient worklist for those dates were seen. A finding written this way cannot be argued with and can be corrected.
After the audit
For each nonconformity the laboratory should:
- correct the immediate problem and consider whether any patient results were affected,
- find the root cause,
- take corrective action that removes the cause,
- check later that the action worked.
Audit results are an input to the management review. Keep the programme, plans, checklists, reports and corrective action records as evidence.
Common weaknesses
- The programme covers the sections but not the management clauses.
- Every audit is a checklist of documents, with no vertical or witness audit.
- Findings are closed with "staff counselled" and no root cause.
- No check that the corrective action worked.
Train as an internal auditor
The ISO 15189 Internal Auditor Course for Laboratory Quality Managers covers the standard clause by clause and the audit method over four study days, with an audit programme, plan, checklist, vertical audit form and nonconformity report you can use. The first lesson is free to read.